A laboratory reflEXAFS spectrometer

Résumé

A simple spectrometer using a conventional X-ray generator has been constructed for ReflEXAFS measurements at glancing angle. The X-ray source is a sealed X-ray tube with a parabolic mirror. A monochromator with two flat LiF crystals is used. The energy resolution amounts to 8.5 eV at 8 keV. A proportional counter monitors the incident X-ray intensity and a NaI scintillation detector measures the intensity of the reflected beam. This system is free from fluctuations of the incident X-ray intensity and harmonics can be rejected with the energy discrimination. The range of energy is 7 to 18 keV. The ReflEXAFS spectra related to the surface of a nickel sample and a thermally oxided nickel sample are presented.